Bolus units, Bolus wizard feature, Carb ratios carb units – Energizer 715 User Manual
Page 166: Daily totals, Default screen

144 Chapter: 11
BG (blood glucose) target
Bolus delivery
Bolus history
Maximum records shown: 24
Bolus units
Increments: 0.1 units
Bolus Wizard feature
(See “Bolus Wizard feature examples” on page 152 for information.)
Carb ratios
Carb units
The food entry ranges when using Bolus Wizard feature:
➠
grams:
0 – 300 (increments: 1 gram)
➠
exchanges:
0.0 – 20 (increments: 0.5 exch)
Daily totals
Maximum records shown: 31 days of data, maximum display: 999.95 units/day. Accuracy: +0/–0.05 units
Default screen
The HOME screen. If no buttons are pressed for 30 seconds (60 seconds for status type screens), the pump returns
to this screen.
Maximum BG targets
range: warning
limits:
8
(mg/dL)
60 – 250
(mmol/L)
3.3 – 13.9
lower limit:
less than 90 mg/dL
upper limit:
greater than 140 mg/dL
Insulin delivered/stroke Fluid delivered/stroke Time between/delivery strokes Delivery rate (per minute)
0.05u
0.5 µL
2 seconds
1.5u
Maximum ratio settings:
range: warning
limits:
8
3 – 150 grams/u
0.1 – 5.0 u/exch
less than 5 or greater than 50 grams/u
less than 0.3 or greater than 3.0 u/exch