Rewinding your pump, Starting on insulin 47 – Energizer 715 User Manual
Page 69

Starting on insulin 47
Rewinding your pump
Before you continue, make sure the pump is NOT connected to your body.
1
If you removed your reservoir and are replacing it, go to the REWIND screen.
Main > Prime > Rewind
2
In the REWIND screen, press
ACT
to start the
rewind process. The REWINDING screen will
appear while the pump rewinds.
3
After the pump rewinds the PREPARING TO PRIME
screen will appear.
NOTE -
If you press
ESC
or the pump returns
to the HOME screen the PRIMING STOPPED
screen will appear which guides you back to the
priming screens. Once you start a rewind, you
cannot cancel it.
If you are practicing, do these steps:
a.
Do NOT insert the reservoir in your pump.
Make sure the shipping cap is installed in the reservoir compartment.
b.
Continue with the manual prime instructions described in “Manual prime” on page 49.
If you are not practicing, continue to “Inserting the reservoir in your pump” on page 48.
WARNING: Make sure the infusion set is disconnected from your body before you rewind or
prime the pump. Never insert the reservoir into the pump while the tubing is
connected to your body. Doing so could result in an accidental infusion of
insulin.
Disconnect from Body
Press ACT to Rewind
Wait for notification
REWIND
REWINDING
ACT
PRIMING STOPPED
Continue Priming
Rewind Pump
Status Screen
PREPARING TO PRIME
Disconnect from Body
Insert/Lock Reservoir
Hold ACT to Prime
ESC
ACT
ACT
time out
(HOME screen)