Pump motor, Dual wave bolus, Easy bolus – Energizer 715 User Manual
Page 167: Infusion pressure, Insulin) sensitivity low resv (reservoir) warning, Pump specifications 145

Pump specifications 145
Pump motor
The pump motor has a unique, patented design with an integrated safety check system. The system delivers in
precise increments.
Dual Wave bolus
Delivers a normal bolus followed by a Square Wave bolus (limited by maximum bolus).
Easy bolus
Features programming using audible tones (or vibrate pulses) in user-defined increments. Beep mode range: 0 to
maximum bolus; Vibrate mode range: 0 to 20 steps or maximum bolus, whichever comes first.
➠
Default step increment:
0.1 unit
➠
Step size
< maximum bolus.
➠
Adjustable step size:
0.1 to 2.0 units per step.
➠
Accessible from the remote control or pump buttons.
Infusion pressure
Maximum infusion pressure and occlusion pressure:
➠
13.7 PSI (pounds per square inch)
➠
94,46 kPa (Kilopascal)
(Insulin) sensitivity
Low resv (reservoir) warning
Values are based on displayed amount, not actual amount.
Maximum settings: factory setting:
range:
warning limits:
8
50 mg/dL (2.8 mmol/L) (mg/dL)
10 – 400 mg/dL
(mmol/L)
0.5 – 22.2 mmol/L
less than 20 or greater than 100 mg/dL
less than 1.1 or greater than 5.6 mmol/L
time:
2 – 24-hours, and @ 1:00 hour before empty
08:00 hours (default when time is selected)
units:
5 – 50 units, and @ 1/2 amount remaining
20 units (factory default)