Pump specifications 153 – Energizer 715 User Manual
Page 175

Pump specifications 153
#3: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 70.
#4: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 100.
#5: Previous bolus activity results in a calculation of 1.5 units unabsorbed (active) insulin. User inputs 60 grams of carbs
and a blood glucose of 200.
70mg/dL - 90mg/dL
40mg/dL/u
-20mg/dL
40mg/dL
+
= 2 + (-0.5)
= 1.5 unit
estimate = 1.5 unit
(correction)
= -0.5 unit
=
60g
30g/u
(food estimate)
= 2 units
+
Correction is 0 because the current blood glucose
reading is between the BG High and Low targets
(correction)
= 2 + 0
= 2 units
estimate = 2 units
60g
30g/u
(food estimate)
= 2 units
60g
30g/u
200mg/dL - 120mg/dL
40mg/dL/u
(food estimate)
(correction)
= 2 units
= 0.5 units
- 1.5 units (active insulin)
= 2 - 1.5
= 2 + 0.5
= 2.5 units
estimate = 2.5 units
+