Bolus wizard feature examples – Energizer 715 User Manual
Page 174

152 Chapter: 11
Bolus Wizard feature examples
Settings:
➠
Carb ratio: 30 grams/unit
➠
Insulin Sensitivity: 40 mg/dL/unit
➠
BG Target: 90-120 mg/dL
➠
Active Insulin Time: 6 Hours
#1: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and does not enter blood glucose.
#2: No active insulin from previous bolus delivery. User inputs 60 grams of carbs and a blood glucose of 200.
estimate = 2 units
60g
30g/u
(food estimate)
= 2 units
200mg/dL - 120mg/dL
40mg/dL/u
= 2 + 2
= 4 units
estimate = 4 units
+
= 2 units
(correction)
60g
30g/u
(food estimate)
= 2 units