Chapter: 9 troubleshooting, My pump has a no delivery alarm, Chapter 9 troubleshooting – Energizer 715 User Manual
Page 151: Troubleshooting, Chapter: 9

Troubleshooting 129
Chapter: 9
Troubleshooting
This chapter is designed to help you understand the messages that your pump is giving you when you
get an alarm or an alert message. The procedures in the beginning of this chapter are to be used when
you get specific alarms or cover a likely condition that might happen. A list of alarms is provided at the
end of this chapter.
NOTE -
It is recommended that you read your warranty statement included with your pump for
information on what is covered during your warranty period.
My pump has a no delivery alarm...
When a “No Delivery” alarm occurs, it means the pump is working correctly. Your pump is not
broken, but it has detected that something is preventing insulin from being delivered. Do the following
steps:
1
Check your blood glucose and take an injection if needed.
2
Make sure that there is insulin in your reservoir and the tubing is not kinked. If these are all right, go
to step
5
.
3
If necessary unkink the tubing. Clear the alarm by pressing
ESC
and
ACT
. A screen will appear with
two choices:
Resume
and
Rewind
. Select
Resume
.
4
If the reservoir is empty, clear the alarm by pressing
ESC
and
ACT
. Select
Rewind
and change your
reservoir and infusion set per the instructions in Chapter 4, “Starting on insulin.”
5
Continue troubleshooting by disconnecting at the quick-disconnect, and set a 10 unit Fixed Prime.
6
Does insulin come out of the needle at the quick-disconnect?
a.
If yes, change your entire infusion set per the instructions in Chapter 4, “Starting on insulin.”
If NO insulin comes out of the needle at the quick-disconnect, or you receive another No
Delivery alarm, call the 24-hour Product Help Line.
b.
Your pump remembers the last fixed prime you deliver, so make sure to set your fixed prime
amount back to your usual setting. To do this, deliver another fixed prime in your normal
amount (the amount specified in your infusion set instructions).