Block feature, Turn block on, Utilities 119 – Energizer 715 User Manual
Page 141

Utilities 119
3
Exit the menus when your are done.
Block feature
The block allows you to prevent (block) people from changing pump setting that should not be
working with the pump s settings. This feature is an important safety feature, especially if the pump
user requires someone else to maintain complete control of pump operation. The factory setting for
this feature is off. When block is on, the remote control is used to deliver a bolus and suspend the
pump. Direct pump programming is limited to suspend, block, and selftest. You can, however, still view
status-type screens (STATUS, BOLUS and PRIME HISTORY, BASAL REVIEW, DAILY TOTALS, etc.). Discuss
what uses and settings are best for you with your healthcare professional. (You can order the remote
control from Medtronic MiniMed.)
Turn block on
NOTE -
The vibrate alert-type is disabled when block is on.
1
Go to the BLOCK OPTION screen.
Main > Utilities > Block
2
Select
On
and press
ACT
. The HOME screen will
appear with an open circle. The Block option is now
on and the pump is in Special mode. Exit the menus.
Add
Delete
Review
Select
Add ID
and press
ACT
.
Select
Delete ID
and press
ACT
.
Select
Review ID
and press
ACT
.
Enter each of the six ID numbers.
Press
ACT
after each.
After you set the last number of
the ID, the screen will return to
the REMOTE ID MENU.
Select the remote ID that you
want to delete and press
ACT
.
The selected ID is now deleted.
The programmed IDs will show
in the REVIEW REMOTE ID
screen.
ADD REMOTE ID
_ _ _ _ _ _
(flashing)
DELETE REMOTE ID
2 222222
3 - - - - - -
1 111111
REMOTE ID MENU
Delete ID
Add ID
Review ID
REVIEW REMOTE ID
1 - - - - -
2 222222
3 - - - - -
BLOCK OPTION
Off
ACT
On