Chapter: 11 pump specifications, Chapter 11 pump specifications, Pump specifications – Energizer 715 User Manual
Page 165: Chapter: 11, Alarms and error messages, Alarm history, Backlight, Basal

Pump specifications 143
Chapter: 11
Pump specifications
This section provides detailed information on specifications related to your pump. The safety features of your
pump are individually listed and described.
Alarms and error messages
➠
Indicators: audible tone (beep) or vibration (silent)
➠
All alarms and errors show messages on the pump’s screen and provide instructions on what to do.
Unresolved alarms will escalate to siren mode for added safety.
Alarm history
➠
The maximum records shown is 36
Backlight
➠
Display type Liquid Crystal Display (LCD)
➠
Time-out period is 30 seconds. The backlight will turn off when you do not push any buttons for 30
seconds or more.
Basal
Maximum patterns
delivery rate
warning limits
3 patterns (with 48 delivery
rates for each pattern)
rate increments:
0.05 units/hour
lowest possible rate:
0.05 units/hour
highest possible rate:
35.0 units/hour
factory maximum rate:
2.0 units/hour
lower limit:
less than 90 mg/dL
upper limit:
greater than 140 mg/dL