Stopping your pump, Bolus and fixed prime). see “stopping your, Suspend – Energizer 715 User Manual
Page 61: Basic programming 39

Basic programming 39
Stopping your pump
You can stop your pump with the suspend function. Suspend stops all insulin delivery including the
current basal and any bolus or prime deliveries that are in progress. While suspended, your pump will
not deliver insulin until you Resume your pump. When basal is resumed, the pump is taken out of the
Suspend mode.
The pump will beep or vibrate about every 15 minutes on the hour to remind you that it is not
delivering insulin. Example: You suspend your pump at 11:20AM. The pump will beep/vibrate at
11:30AM, 11:45AM, 12:00PM, and so on until you resume your pump (basal resumes).
NOTE -
When suspended, your pump is in Attention mode (a solid circle). When in Suspend, you can
only resume your basal or view the STATUS screen. No other functions are available.
Do these steps to suspend your pump:
NOTE -
Press
ESC
to view the STATUS screen
and verify your pump is suspended.
1
Select
Suspend
from
the MAIN MENU, and
press
ACT
.
2
SUSPEND
will flash. Press
ACT
to stop your pump.
3
The screen will show that the pump
is suspended and the time that it
stopped. After a few minutes, the
pump will default to the HOME
screen with a solid circle.
MAIN MENU
Bolus
Suspend
Basal
ACT
(flashing)
SUSPEND
Press ACT to Stop Pump
ACT
Stopped at 1:05P
SUSPEND
STATUS U100
Suspended at 01:05P
Last Bolus N 3.8U
7:58A 22 FEB