12 Chapter: 2
Install battery
Medtronic MiniMed designed the pump to only accept a NEW battery. As a safety measure, if you
install a battery that does not have full power, the “weak battery” or “failed batt test” alarms may
sound. If a “weak battery” alarm happens, respond to the alarm and continue. The pump will still
operate normally, but with a decreased battery life. The pump uses one AAA alkaline battery.
1
Make sure all the following apply:
➠
Clear (
ESC
,
ACT
) any alarms and/or alerts before removing and replacing the battery
➠
Make sure the pump is at the HOME (idle) screen when you remove the battery.
➠
Do NOT remove the battery during a bolus or prime delivery.
2
Use the edge of a coin to remove the battery cap. Turn the cap in a counter-clockwise direction.
3
Remove the old battery and dispose of it per the
disposable requirements of your state or country. Put
the new battery in the pump with the negative end
[(-) symbol] going in first. Check the label on the back
of the pump to make sure the battery is inserted
correctly.
NOTE -
Do not use batteries that have been in cold storage (i.e., in the refrigerator or your car in
the winter).
4
Place the battery cap in the pump and tighten so the slot is
aligned horizontally with the pump as shown here:
Do NOT apply force when you turn the cap.
CAUTION:
Do not use a rechargeable or carbon zinc battery in your pump. For best
results, a new AAA Energizer alkaline battery is recommended.
CAUTION:
Do NOT over-tighten the battery cap.
You should not turn the cap more than
four half turns. If you over-tighten the
cap you may not be able to remove it,
and you can damage your pump.
Battery cap
Battery
one half-turn
one half-turn