Seaward DO7010 User Manual
Page 39
![background image](https://www.manualsdir.com/files/804147/content/doc039.png)
5-3-3 Example for a FAIL measurement:
Φ Trigger a new measurement.
Φ If the value is higher than the limit, an audible signal (2 beeps) sounds, the red
LED lights, and the DO7010 unit stays on the same test step.
Φ To re-test, trigger a new measurement either with the probe contact or the
green front panel push button.
Φ To store a FAIL measurement, press on the VALID function key.
Φ Confirm the validation by pressing on the YES function key.
Φ Then the DO7010 unit jumps to the next test step.
Φ To cancel the FAIL confirmation, press on the NO function key and trigger a
new measurement.
5-3-4 Example for an ABSENT point:
Φ If a point to be measured is absent or non available for the measurement, then
press on the ABSENT function key.
Φ Confirm the ABSENT status by pressing on the YES function key.
Φ Then the unit jumps to the next test step.
Φ To cancel the ABSENT confirmation, press on the NO function key and
trigger a new measurement.
ABCDE123456789
S:500.0mΩ
Ι:10Α
BAT:80%
154: ABS
mΩ
POINT ABSENT
VALID
SET UP
ESC
ABCDE123456789
S:10.00mΩ
Ι:10Α
BAT:80%
153: 14.25
mΩ
POINT ABSENT
VALID
SET UP
ESC
Operating Instructions
Digital Microhmmeter Type DO7010
27
•
•
•
•
•
•
•
•
•
•
•