Putting in contact 2 metals from different material generates unwanted voltages
(EMF) proportional to the ambient temperature.

Two methods exist to cancel this problem: the Zero method and the Average method.

The ZERO method:

Φ Select the ZERO line

ON

= Zero method operating

OFF

= Zero method not operating

Φ Enter the selection with the Enter key.

The AVERAGE method:

Φ Select the AVERAGE line

Φ Press on the RIGHT arrow or on the enter key.

measurements to get the resistance calculation. Select with the UP and DOWN
arrows:

ON

= Average method operating

OFF

= Average method not operating

Φ Enter the selection with the Enter key.

RANGE OFF

CURRENT

ON

ZERO

AVERAGE

>

LOG

LIMITS

ESC

RANGE OFF

CURRENT

ON

ZERO

>

AVERAGE

LOG

LIMITS

ESC

When measuring low resistances, consequently low voltages, the accessories and

4-1-3 Thermal EMF Error:

Both methods are available with the DO7010.

Φ The Zero method consists of a voltage measurement without current, then a

Φ The Average method consists of a voltage measurement with a positive

resistance calculation. Select with the UP and DOWN arrows:

Φ Press on the RIGHT arrow or on the enter key.

voltage measurement with current, and subtraction of EMF voltages before

current, then a voltage measurement with negative current, and average of the 2

specimen under test conductive metals can introduce significant errors.

Operating Instructions

Digital Microhmmeter Type DO7010

28