1-2-10 Remote Control Unit (RCU) specifications (DO7010-REM)

On option, the DO7010 unit can work in conjunction with a remote control unit
(DO7010-REM)
The DO7010 RCU is made in a shock proof black colour polystyrene case. It includes
a 64x240 dot matrix graphic LCD screen with backlight, a beeper, red and green LED,
a set of 5 function keys and a 4 direction navigator with an enter key.
The RCU also includes a battery back up memory, allowing uploading measurement
results in to a PC computer without having to transport the DO7010 main unit.
The RCU is connected to the DO7010 with a 3 meter long cable, which can be
extended to 15 meters.

RCU dimensions:

- Length

: 200 mm

- Width

: 120 mm

- High

: 60 mm

Weight:

- 600 g

Display:

64x240 points LCD matrix (liquid crystal display) with electroluminescent
diode backlight.
6000 digit measurement and location test point display in large and small
characters for:
-
-

measurement range

-

measurement current

-

battery charge level

-

Error code (measurement, polarity…)

Keyboard:

-

5 function keys changing according to the menus.

-

4 navigation keys: arrows UP, DOWN, RIGHT, LEFT.

-

1 data enter key

-

1 software power ON key

-

1 mechanical key to trigger the measurement.

Visual and audible signal:

-

Audible signal when pressing keyboard ( can be disabled)

-

Audible signal with error message and FAILED measurements

-

Red and Green LEDS for PASS/FAIL measurement results.

Battery charge level:

From 0 to 99%

Operation:

The RCU can be operated either by hand, or strapped to the fore arm to make
the measurement easier.

units

Operating Instructions

Digital Microhmmeter Type DO7010

10