4-2 Measurements in MANUAL mode:

Φ Trigger a measurement with the green push button on the battery pack front

panel, or by making contact on the specimen under test with the probes (see
trigger modes in section 3-6)

Φ If measurement probes or leads are not correctly connected to the specimen

under test, the following error messages can be displayed :

“CIRCUIT INCOMPLETE CHECK +VE CABLES”

“CIRCUIT INCOMPLETE CHECK -VE CABLES”

4-3 Measurements in MANUAL mode with the limit function:

Φ Set the LIMITS mode (see chapter 4-1-5) and trigger a measurement with the

green push button on the battery pack front panel, or by making contact on the
specimen under test with the probes.

Φ If the measurement result is :

Within the limits, the display shows

: OK

Lower than the low limit, the display shows

: LO

Higher than the high limit, the display shows

: HI

RANGE:600mΩ

I:10A (HI)

ΒΑΤ:80%

>>OK<<

100.2mΩ

NO FUEL

PARAM

SETUP

ESC

RANGE:600mΩ

I:10A (HI)

ΒΑΤ:80%

NO FUEL

100.2mΩ

PARAM

SETUP

ESC

“RESISTANCE TOO HIGH IN CURRENT LEADS”

Operating Instructions

Digital Microhmmeter Type DO7010

31