4-4 Measurement in MANUAL mode with the log mode:

WARNING: this feature is only available if there is no Sequence downloaded in the
DO7010 (see chapter 5).To delete the loaded sequence; please refer to the DO7010-
PRO software manual.

Φ Set the Log mode (see section 4-1-4) and trigger a measurement with the green

push button on the battery pack front panel, or by making contact on the
specimen under test with the probes.

Φ Each of the measurements will be stored together with the range, current, date,

and time.

Φ To review the measured values:

Press on the PARAM function key

Select the LOG line

Press on the Enter key

Select the REVIEW line

Press on the Enter key

The LCD screen shows the values as follows:

-

Scroll the values with the DOWN or UP arrows.

-

Escape this menu with the ESC function key.

Press on the PARAM function key

Select the LOG line

Press on the Enter key

Select the DELETE line

Press on the Enter key

Confirm the delete by selecting the CONFIRM line

Press on the Enter key.

Select the CANCEL line to stop delete operation

Press on the Enter key

LOG#

Rx RANGE

I DATE

TIME

1 10.01

60 HI 19/07/04 13 :14

2 10.03

60 HI 19/07/04 13 :15

3 10.01

60 HI 19/07/04 13 :16

4 10.05

60 HI 19/07/04 13 :17

5 10.01

60 HI 19/07/04 13 :18

To delete all the saved measurements:

Operating Instructions

Digital Microhmmeter Type DO7010

32