SECTION 4: OPERATING THE DO7010 IN MANUAL MODE
In addition the DO7010 SEQUENCE operating mode performs measurement
sequence according to saved protocol through PC download (see section 5)
MANUAL menu.
Φ The LCD screen upper line gives range and current parameters.
Φ The
NO FUEL
message is displayed (according to manufacture setup) for
Φ To change the measurement ranges, press on the UP or DOWN arrows.
Φ To change the other parameters, press on the PARAM function key.
Φ With the UP and DOWN arrows, move the reverse video area to select the line
to modify.
Φ With the RIGHT arrow or the enter key, enter in the menu level to set the
parameters.
RANGE
>
600mΩ
CURRENT HI
ZERO
OFF
AVERAGE OFF
LOG OFF
LIMITS
ON
ESC
RANGE:600mΩ
I:10A (HI)
ΒΑΤ:80%
NO FUEL
- - - - mΩ
PARAM
SETUP
ESC
The DO7010 MANUAL operating mode allows single measurement as any
From the power ON screen, press on the MANUAL function key to reach the
ordinary measurement unit.
Φ After each power ON, those parameters will be reset to default values (60mΩ,
current values :1A and 10A
10A, ZERO mode ) or memorized (according to manufacture setup).
4-1 MANUAL mode parameters:
Operating Instructions
Digital Microhmmeter Type DO7010
26
•
•
•
•
•
•
•