4-1-1 Measurement range selection:

Φ Select the RANGE line

Φ Press on the RIGHT arrow or on the enter key .

6000m

Ω : from 0001 to 6000 mΩ

600m

Ω : from 000.1 to 600.0 mΩ

60m

Ω : from 00.01 to 60.00 mΩ

6m

Ω : from 0.001 to 6.000 mΩ

Φ Enter the selection with the Enter key.

4-1-2 Measurement current selection:

Φ Select the CURRENT line

Φ Press on the RIGHT arrow or on the enter key.

Φ Select the measurement current with the UP and DOWN arrows:

HI = 10A for the 6 to 600m

Ω ranges and 1A for the 6000 mΩ range.

LO = 1A for the 6 to 600m

Ω and 0.1A for the 6000 mΩ range.

Φ Enter the selection with the Enter key.

RANGE LO

CURRENT

>

HI

ZERO

AVERAGE

LOG

LIMITS

ESC

RANGE

>

6000mΩ

CURRENT 600mΩ

ZERO

60mΩ

AVERAGE 6mΩ

LOG AUTO1

LIMITS

AUTO2

ESC

Φ Select the measurement range with the UP and DOWN arrows:

AUTO2 : automatic searching range from the last measurement range.

AUTO1 : automatic searching range from the highest range (6000m ).

Ω

Operating Instructions

Digital Microhmmeter Type DO7010