1-4 MEASUREMENT METHOD:

In order to perform high accuracy low resistance measurements, the DO7010 uses a 4
wire method named Kelvin method.

-

2 wires are used to connect a constant current source to the specimen under
test

-

2 wires are used to measure the voltage difference directly on the specimen
under test terminals.

With this method the measurement leads resistance doesn’t introduce error, but should
not exceed 170 m

Ω for the current connection.

Kelvin method

-I

R lead

R lead

R lead

Rx

Ic

+I

+U

-U

R lead

Ux

Rx = Ux / Ic

DO7010

Accessories

Resistance to
be measured

Operating Instructions

Digital Microhmmeter Type DO7010