4-

4

Windows and Menus

878493

eZ-NDT

Digital I/O Panel

Digital I/O Panel

The Digital I/O Panel is used with the optional digital I/O control package. It is associated with the
Configuration window’s Automation tab, which is discussed later in this chapter

The Digital I/O panel’s toggle switch controls the ON/OFF status of the eZ-NDT Automation Module. The
four Relay Icons indicate the current status of the relays.

Each time you start the NDT software, the Automation main switch is turned “off.” You must turn the
switch “on” to send output to the relay channels.

The relay channels can be set to trigger according to:

the state of a channel

the number of consecutive fails

the acquisition status

You can test a relay by clicking on its associated button.

Note that if Digital Inputs have been enabled, the digital input status will be present in the label above the
output relay buttons.