Digital i/o panel – Measurement Computing eZ-NDT version 6.0.9 User Manual
Page 24

4-
4
Windows and Menus
878493
eZ-NDT
Digital I/O Panel
Digital I/O Panel
The Digital I/O Panel is used with the optional digital I/O control package. It is associated with the
Configuration window’s Automation tab, which is discussed later in this chapter
The Digital I/O panel’s toggle switch controls the ON/OFF status of the eZ-NDT Automation Module. The
four Relay Icons indicate the current status of the relays.
Each time you start the NDT software, the Automation main switch is turned “off.” You must turn the
switch “on” to send output to the relay channels.
The relay channels can be set to trigger according to:
•
the state of a channel
•
the number of consecutive fails
•
the acquisition status
You can test a relay by clicking on its associated button.
Note that if Digital Inputs have been enabled, the digital input status will be present in the label above the
output relay buttons.