Counters/timers– adac/5500mf, Physical & environmental - adac/5500mf – Measurement Computing ADAC/5500 Series User Manual
Page 47

6.1.4 Counters/Timers–
ADAC/5500MF
Number:
2 Counters / 2 Timers
Counter 0
Shared with ADCLKIN terminal
Counter 1
Shared with DACLKIN terminal
Counter 0/1Input Delay:
- 100ns uncertainty
Counter 0/1 Maximum Count
65536
Counter 0/1 Maximum Input Freq.
900Khz
Timer 0
Shared with ADTGOUT terminal
Timer 1
Shared with ADCLKOUT terminal
Timer 0/1 Output Rate
7.6Htz to 500Khz 50 % Duty Cycle Square Wave
Counter Input Levels:
5 V CMOS/TTL with 4.7 Kohm pull-up resistor
High Level Input Voltage:
2 V min.
Low Level Input Voltage:
0.8 V max.
Timer Output Levels:
5 V CMOS/TTL with 4.7 Kohm pull-up resistor
High Level Output Voltage:
2.4 V
Low Level Output Voltage:
0.5 V
Maximum Output Current:
Low: 24 mA (sinking)
High: 24 mA (sourcing)
6.1.5
Physical & Environmental - ADAC/5500MF
Size:
4.2" (10.6cm) H x 5.57" (14.1cm) L
Connector:
68 Pin standard SCSI Type III Female Connector brought
to outside of PC.
Temp. Range of Operation:
0 °C – 55 °C
CE Conformity:
EN 55022 Class B
EN50082-1
IEC801-2
IEC801-3
IEC801-4
ADAC Series
908196
-42-
ADAC/5500 Series User’s Manual