SECTION 2

CALIBRATION

2.1 CALIBRATION

SETUP

Prior to starting the calibration procedure, gather the necessary equipment
(Section 2.1.1), then power up the chassis and allow it to warm up for 15
minutes.

2.1.1

E

QUIPMENT

R

EQUIRED

DC Voltage Standard: Krohn-Hite EDC 521

Current Shunt: Prodigit 7550

Digital Multi-Meter (DMM): Agilent 34401A

DC Power Supply: Sorensen SG Series (Supply depends upon
models being calibrated.)

2.2

DVM CALIBRATION

Connect the SL Series electronic load as shown in Figure 2-1

Figure 2-1. Connections for DVM Calibration Items 1-4)

Press MODE and LEVEL keys simultaneously to enter Calibration Mode and
to be at Item 1 to start calibration.

SLM/SLH Calibration Manual

2-1