AMETEK SL Series Calibration Manual User Manual
Page 29
Elgar Electronics Corporation
Calibration
2.5.3
CR M
ODE
, L
EVEL
L
OW
, R
ANGE
II O
FFSET AND
G
AIN
, I
TEMS
25
AND
26
1.
LOAD OFF.
2.
Press the
© key to go to Item 25.
3.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-25 for the model being
calibrated.
4.
PRES ON.
5.
Set the current to the Current Setting value in Table 2-25 for the
model being calibrated.
6.
PRES OFF.
7.
LOAD ON.
8.
Adjust using the
keys until the DAM reading and the current
setting match.
MODEL Resistance
Setting
VSENSE Input
Current Setting
SLM-60-30-150 0.1068Ω 3.204V 30.00A
SLM-60-60-300 0.0534Ω 3.204V 60.00A
SLM-250-10-300 1.3334Ω 13.334V 10.00A
SLM-500-10-300 2.680Ω 26.80V 10.000A
SLM-60-15-75 0.2134Ω 3.201V 15.00A
SLH-60-120-600 0.0268
Ω 3.216V 120.00A
SLH-60-120-1200 0.0268
Ω 3.216V 120.00A
SLH-60-120-1800 0.0268
Ω 3.216V 120.00A
SLH-60-240-1200 0.0133
Ω 3.216V 240.00A
SLH-60-240-1800 0.0134
Ω 3.216V 240.00A
SLH-60-360-1800 0.0088
Ω 3.168V 360.00A
SLH-500-60-1800 0.4444Ω 26.666V 60.00A
Table 2-25. CR Mode, Level Low, Range II Offset Adjustment Settings (Item 25)
9.
LOAD OFF.
10.
Press the
© key to go to Item 26.
SLM/SLH Calibration Manual
2-21