beautypg.com

AMETEK SL Series Calibration Manual User Manual

Page 26

background image

Calibration

Elgar Electronics Corporation

11.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-22 for the model being
calibrated.

12.

PRES ON.

13.

Set the current to the Current Setting value in Table 2-22 for the
model being calibrated.

14.

PRES OFF.

15.

LOAD ON.

16.

Adjust using the

…† keys until the DAM reading and the current

setting match.

17.

Repeat steps 1 through 16 (press

ª to return to Item 21) until no

further adjustments are necessary for the DAM reading t and the
current setting to match.

MODEL Resistance

Setting

VSENSE Input

Current Setting

SLM-60-30-150 2.000

Ω 60.00V 30.00A

SLM-60-60-300 1.000Ω 60.00V 60.00A
SLM-250-10-300 25.00Ω 250.00V 10.00A
SLM-500-10-300 50.00Ω 500.00V 10.000A
SLM-60-15-75 4.000Ω 60.00V 15.000A

SLH-60-120-600 0.5000

Ω 60.00V 120.00A

SLH-60-120-1200 0.5000

Ω 60.00V 120.00A

SLH-60-120-1800 0.5000

Ω 60.00V 120.00A

SLH-60-240-1200 0.2500

Ω 60.00V 240.00A

SLH-60-240-1800 0.2500

Ω 60.00V 240.00A

SLH-60-360-1800 0.167

Ω 60.00V 359.281A

SLH-500-60-1800 8.3333

Ω 500.00V 60.00A

Table 2-22. CR Mode, Level Low, Range I Gain Adjustment Settings (Item 22)

2-18

SLM/SLH Calibration Manual