AMETEK SL Series Calibration Manual User Manual
Page 26
Calibration
Elgar Electronics Corporation
11.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-22 for the model being
calibrated.
12.
PRES ON.
13.
Set the current to the Current Setting value in Table 2-22 for the
model being calibrated.
14.
PRES OFF.
15.
LOAD ON.
16.
Adjust using the
keys until the DAM reading and the current
setting match.
17.
Repeat steps 1 through 16 (press
ª to return to Item 21) until no
further adjustments are necessary for the DAM reading t and the
current setting to match.
MODEL Resistance
Setting
VSENSE Input
Current Setting
SLM-60-30-150 2.000
Ω 60.00V 30.00A
SLM-60-60-300 1.000Ω 60.00V 60.00A
SLM-250-10-300 25.00Ω 250.00V 10.00A
SLM-500-10-300 50.00Ω 500.00V 10.000A
SLM-60-15-75 4.000Ω 60.00V 15.000A
SLH-60-120-600 0.5000
Ω 60.00V 120.00A
SLH-60-120-1200 0.5000
Ω 60.00V 120.00A
SLH-60-120-1800 0.5000
Ω 60.00V 120.00A
SLH-60-240-1200 0.2500
Ω 60.00V 240.00A
SLH-60-240-1800 0.2500
Ω 60.00V 240.00A
SLH-60-360-1800 0.167
Ω 60.00V 359.281A
SLH-500-60-1800 8.3333
Ω 500.00V 60.00A
Table 2-22. CR Mode, Level Low, Range I Gain Adjustment Settings (Item 22)
2-18
SLM/SLH Calibration Manual