Calibration

Elgar Electronics Corporation

8.

LOAD OFF.

9.

Press the

© key to go to Item 10.

10.

PRES ON.

11.

Set the load current as shown in Table 2-10 for the model being
calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the shunt current and the load DAM

reading match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 9) until no

further adjustments are necessary for the load DAM to match the
DMM reading.

MODEL Current

Setting

MODEL Current

Setting

SLM-60-30-150 3.000A SLH-60-120-600 12.00A
SLM-60-60-300 6.000A SLH-60-120-1200

12.00A

SLM-250-10-300 1.0050A SLH-60-120-1800 12.00A
SLM-500-10-300 1.0050A SLH-60-240-1200 24.00A
SLM-60-15-75 1.500A

SLH-60-240-1800

24.00A

SLH-60-360-1800

36.00A

SLH-500-60-1800

6.000A

Table 2-10. DAM, Range I Gain Adjustment Settings (Item 10)

2-8

SLM/SLH Calibration Manual