Calibration

2.4.2

DAM, R

ANGE

II O

FFSET AND

G

AIN

, I

TEMS

11

AND

12

1.

LOAD OFF.

2.

Press the

© key to go to Item 11.

3.

PRES ON.

4.

Set the load current as shown in Table 2-11 for the model being
calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the load DAM matches the DMM

reading.

MODEL Current

Setting

MODEL Current

Setting

SLM-60-30-150 0.00A SLH-60-120-600 0.00A
SLM-60-60-300 0.00A SLH-60-120-1200

0.00A

SLM-250-10-300 0.00A SLH-60-120-1800 0.00A
SLM-500-10-300 0.00A SLH-60-240-1200 0.00A
SLM-60-15-75 0.00A

SLH-60-240-1800

0.00A

SLH-60-360-1800

0.00A

SLH-500-60-1800

0.00A

Table 2-11. DAM, Range II Offset Adjustment Settings (Item 11)

8.

Load OFF.

9.

Press the

© key to go to Item 12.

SLM/SLH Calibration Manual

2-9