AMETEK SL Series Calibration Manual User Manual
Page 20
Calibration
Elgar Electronics Corporation
2.4.4
CC M
ODE
, L
EVEL
H
IGH
, R
ANGE
I O
FFSET AND
G
AIN
, I
TEMS
15
AND
16
1.
LOAD OFF.
2.
Press the
© key to go to Item 15.
3.
PRES ON.
4.
Set the load current as shown in Table 2-15 for the model being
calibrated.
5.
PRES OFF.
6.
LOAD ON.
7.
Adjust using the
keys until the shunt current and the current
setting match.
MODEL
Current
Setting MODEL Current
Setting
SLM-60-30-150 0.00A
SLH-60-120-600 0.00A
SLM-60-60-300 0.00A
SLH-60-120-1200 0.00A
SLM-250-10-300 0.00A
SLH-60-120-1800 0.00A
SLM-500-10-300 0.00A
SLH-60-240-1200 0.00A
SLM-60-15-75 0.00A
SLH-60-240-1800
0.00A
SLH-60-360-1800
0.00A
SLH-500-60-1800
0.00A
Table 2-15. CC Mode, Level High, Range I Offset Adjustment Settings (Item 15)
8.
LOAD OFF.
9.
Press the
© key to go to Item 16.
10.
PRES ON.
11.
Set the load current as shown in Table 2-16 for the model being
calibrated.
12.
PRES OFF.
13.
LOAD ON.
14.
Adjust using the
keys until the shunt current and the current
setting match.
15.
Repeat steps 1 through 14 (press
ª to return to Item 15) until no
further adjustments are necessary for the shunt current and the
current setting to match.
2-12
SLM/SLH Calibration Manual