AMETEK SL Series Calibration Manual User Manual
Page 25

Elgar Electronics Corporation
Calibration
2.5.1
CR M
ODE
, L
EVEL
L
OW
, R
ANGE
I O
FFSET AND
G
AIN
, I
TEMS
21
AND
22
1.
LOAD OFF.
2.
Press the
© key to go to Item 21.
3.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-21 for the model being
calibrated.
4.
PRES ON
5.
Set the current to the Current Setting value in Table 2-21 for the
model being calibrated.
6.
PRES OFF.
7.
LOAD ON.
8.
Adjust using the
keys until the DAM reading and the current
setting match.
MODEL Resistance
Setting
VSENSE Input
Current Setting
SLM-60-30-150 3750
Ω 60.00V 0.016A
SLM-60-60-300 3750
Ω 60.00V 0.016A
SLM-250-10-300 15625
Ω 250.00V 0.016A
SLM-500-10-300 6250
Ω 500.0V 0.08A
SLM-60-15-75 5000
Ω 60.00V 0.012A
SLH-60-120-600 1875
Ω 60.00V 0.0320A
SLH-60-120-1200 1875
Ω 60.00V 0.0320A
SLH-60-120-1800 1875
Ω 60.00V 0.0320A
SLH-60-240-1200 937.50
Ω 60.00V 0.0640A
SLH-60-240-1800 937.50
Ω 60.00V 0.0640A
SLH-60-360-1800 625
Ω 60.00V 0.0960A
SLH-500-60-1800 15625.
Ω 500.00V 0.0320A
Table 2-21. CR Mode, Level Low, Range I Offset Adjustment Settings (Item 21)
9.
LOAD OFF.
10.
Press the
© key to go to Item 22.
SLM/SLH Calibration Manual
2-17