AMETEK SL Series Calibration Manual User Manual
Page 21

Elgar Electronics Corporation
Calibration
16.
PRES ON.
17.
Set the load current as shown in Table 2-16 for the model being
calibrated.
18.
PRES OFF.
19.
LOAD ON.
20.
Adjust using the
keys until the shunt current and the current
setting match.
21.
Repeat steps 1 through 14 (press
ª to return to Item 15) until no
further adjustments are necessary for the shunt current and the
current setting to match.
Current
Setting
MODEL
MODEL Current
Setting
SLM-60-30-150 3.000A
SLH-60-120-600 12.00A
SLM-60-60-300 6.000A
SLH-60-120-1200
12.00A
SLM-250-10-300 1.0050A
SLH-60-120-1800
12.00A
SLM-500-10-300 1.0050A
SLH-60-240-1200
24.00A
SLM-60-15-75 1.500A
SLH-60-240-1800
24.00A
SLH-60-360-1800
36.00A
SLH-500-60-1800
6.000A
Table 2-16. CC Mode, Level High, Range I Gain Adjustment Settings (Item 16)
2.4.5
CC M
ODE
, L
EVEL
L
OW
, R
ANGE
II O
FFSET AND
G
AIN
, I
TEMS
17
AND
18
1.
LOAD OFF.
2.
Press the
© key to go to Item 17.
3.
PRES ON.
4.
Set the load current as shown in Table 2-17 for the model being
calibrated.
5.
PRES OFF.
6.
LOAD ON.
7.
Adjust using the
keys until the shunt current and the current
setting match.
MODEL
Current Setting
MODEL
Current Setting
SLM-60-30-150 0.00A
SLH-60-120-600 0.00A
SLM-60-60-300 0.00A
SLH-60-120-1200 0.00A
SLM-250-10-300 0.00A
SLH-60-120-1800 0.00A
SLM-500-10-300 0.00A
SLH-60-240-1200 0.00A
SLM-60-15-75 0.00A
SLH-60-360-1800
0.00A
SLH-500-60-1800
0.00A
Table 2-17. CC Mode, Level Low, Range II Offset Adjustment Settings (Item 17)
SLM/SLH Calibration Manual
2-13