Calibration
Elgar Electronics Corporation
11.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-28 for the model being
calibrated.
12.
PRES ON.
13.
Set the current to the Current Setting value in Table 2-28 for the
model being calibrated.
14.
PRES OFF.
15.
LOAD ON.
16.
Adjust using the
keys until the DAM reading and the current
setting match.
17.
Repeat steps 1 through 16 (press
ª to return to Item 27) until no
further adjustments are necessary for the DAM reading and the
current setting to match.
MODEL
Resistance Setting
VSENSE
Current Setting
SLM-60-30-150 2.000Ω 60.00V
30.000A
SLM-60-60-300 1.0001Ω 60.00V
60.000A
SLM-250-10-300 25.00Ω 250.00V
10.000A
SLM-500-10-300 50.000Ω 500.00V
10.000A
SLM-60-15-75 4.000Ω 60.00V
15.000A
SLH-60-120-600 0.5000
Ω 60.00V
120.00A
SLH-60-120-1200 0.5000
Ω 60.00V
120.00A
SLH-60-120-1800 0.5000
Ω 60.00V
120.00A
SLH-60-240-1200 0.2500
Ω 60.00V
240.00A
SLH-60-240-1800 0.2500
Ω 60.00V
240.00A
SLH-60-360-1800 0.167
Ω 60.00V
359.281A
SLH-500-60-1800 8.3333
Ω 500.00V 60.00A
Table 2-28. CR Mode, Level High, Range II Gain Adjustment Settings (Item 28)
2-24
SLM/SLH Calibration Manual