beautypg.com

AMETEK SL Series Calibration Manual User Manual

Page 22

background image

Calibration

Elgar Electronics Corporation

8.

LOAD OFF.

9.

Press the

© key to go to Item 18.

10.

PRES ON.

11.

Set the load current as shown in Table 2-18 for the model being
calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the shunt current and the current

setting match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 17) until no

further adjustments are necessary for the shunt and the current
setting to match.

MODEL

Current Setting

MODEL

Current Setting

SLM-60-30-150 30.00A

SLH-60-120-600

120.00A

SLM-60-60-300 60.00A

SLH-60-120-1200

120.00A

SLM-250-10-300 10.050A

SLH-60-120-1800

120.00A

SLM-500-10-300 10.050A

SLH-60-240-1200

240.00A

SLM-60-15-75 15.000A

SLH-60-240-1800

240.00A

SLH-60-360-1800 360.00A

SLH-500-60-1800 60.00A

Table 2-18. CC Mode, Level Low, Range II Gain Adjustment Settings (Item 18)

2-14

SLM/SLH Calibration Manual