AMETEK SL Series Calibration Manual User Manual
Page 31

Elgar Electronics Corporation
Calibration
2.5.4
CR M
ODE
, L
EVEL
H
IGH
, R
ANGE
II O
FFSET AND
G
AIN
, I
TEMS
27
AND
28
1.
LOAD OFF.
2.
Press the
© key to go to Item 27.
3.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-27 for the model being
calibrated.
4.
PRES ON.
5.
Set the current to the Current Setting value in Table 2-27 for the
model being calibrated.
6.
PRES OFF.
7.
LOAD ON.
8.
Adjust using the
keys until the DAM reading and the current
setting match.
MODEL
Resistance Setting VSENSE Input Current Setting
SLM-60-30-150 0.1068Ω 3.204V
30.000A
SLM-60-60-300 0.0534Ω 3.204V
60.000A
SLM-250-10-300 1.3334Ω 13.334V
10.000A
SLM-500-10-300 2.680Ω 26.80V
5.000A
SLM-60-15-75 0.2134Ω 3.201V
15.000A
SLH-60-120-600 0.0268
Ω 3.216V
120.00A
SLH-60-120-1200 0.0268
Ω 3.216V
120.00A
SLH-60-120-1800 0.0268
Ω 3.216V
120.00A
SLH-60-240-1200 0.0133
Ω 3.216V
240.00A
SLH-60-240-1800 0.0134
Ω 3.216V
240.00A
SLH-60-360-1800 0.0088
Ω 3.168V
360.00A
SLH-500-60-1800 0.4444Ω 26.666V 60.00A
Table 2-27. CR Mode, Level High, Range II Offset Adjustment Settings (Item 27)
9.
LOAD OFF.
10.
Press the
© key to go to Item 28.
SLM/SLH Calibration Manual
2-23