AMETEK SL Series Calibration Manual User Manual
Page 27
Elgar Electronics Corporation
Calibration
2.5.2
CR M
ODE
, L
EVEL
H
IGH
, R
ANGE
I O
FFSET AND
G
AIN
I
TEMS
23
AND
24
1.
LOAD OFF.
2.
Press the
© key to go to Item 23.
3.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-23 for the model being
calibrated.
4.
PRES ON.
5.
Set the current to the Current Setting value in Table 2-23 for the
model being calibrated.
6.
PRES OFF.
7.
LOAD ON.
8.
Adjust using the
keys until the DAM reading and the current
setting match.
MODEL Resistance
Setting
VSENSE Input Current Setting
SLM-60-30-150 3750
Ω 60.00V
0.016A
SLM-60-60-300 3750
Ω 60.00V
0.016A
SLM-250-10-300 15625
Ω 250.00V 0.08A
SLM-500-10-300 6250
Ω 500.0V 0.04A
SLM-60-15-75 5000
Ω 60.00V
0.012A
SLH-60-120-600 1875
Ω 60.00V
0.0320A
SLH-60-120-1200 1875
Ω 60.00V
0.0320A
SLH-60-120-1800 1875
Ω 60.00V
0.0320A
SLH-60-240-1200 937.50
Ω 60.00V
0.0640A
SLH-60-240-1800 937.50
Ω 60.00V
0.0640A
SLH-60-360-1800 625
Ω 60.00V
0.0960A
SLH-500-60-1800 15625.
Ω 500.00V 0.032A
Table 2-23. CR Mode, Level High, Range I Offset Adjustment Settings (Item 23)
9.
LOAD OFF.
10.
Press the
© key to go to Item 24.
SLM/SLH Calibration Manual
2-19