8 store calibration data, Tore, Alibration – AMETEK SL Series Calibration Manual User Manual
Page 41
Elgar Electronics Corporation
Calibration
9.
Input Voltage to the VSENSE terminal as Table 2-36.
10.
PRES ON, setting Power as table 40, PRES OFF.
11.
LOAD ON, Adjust using the
keys until load DAM matches
Current Setting value.
12.
Repeat steps 1 through 11 (press
ª to return to Item 21) until no
further adjustments are necessary for the load DAM to match the
DMM reading.
13.
LOAD OFF.
MODEL
Power Setting
VSENSE Input
Current Setting
SLM-60-30-150 150W
5.00V
30.000A
SLM-60-60-300 300W 5.00V 60.000A
SLM-250-10-300 300W 30.00V 10.000A
SLM-500-10-300 300W 30.00V 10.000A
SLM-60-15-75 75W 5.00V
15.000A
SLH-60-120-600 600.0W 5.000V 120.0A
SLH-60-120-1200 1200.0W 10.000V 120.0A
SLH-60-120-1800 1800.0W 15.000V 120.0A
SLH-60-240-1200 1200.0W 5.000V
240.0A
SLH-60-240-1800 1800.0W 7.500V
240.0A
SLH-60-360-1800 1800.0W 5.000V
360.0A
SLH-500-60-1800 1800.0W 30.00V
60.00A
Table 2-36. CP Mode, Level High, Gain Adjustment Settings (Item 36)
2.8
STORE CALIBRATION DATA
Press SHORT key to store the calibration data.
Press LIMIT key to exit the calibration mode.
SLM/SLH Calibration Manual
2-33