beautypg.com

AMETEK SL Series Calibration Manual User Manual

Page 14

background image

Calibration

Elgar Electronics Corporation

5.

Set the DC Voltage Standard to the Input Voltage value shown in Table
2-8
for the model being calibrated.

6.

Use the

…† keys to adjust the load DVM display until it matches the

DMM reading.

7.

Repeat steps 1 through 6 (press

ª to return to Item 7) until no further

adjustments are necessary for the DVM display to match the DMM
reading.

MODEL Input

Voltage

MODEL Input

Voltage

SLM-60-30-150 60.00V SLH-60-120-600 60.00V
SLM-60-60-300 60.00V SLH-60-120-1200 60.00V
SLM-250-10-300 250.00V SLH-60-120-1800 60.00V
SLM-500-10-300 500.00V SLH-60-240-1200 60.00V
SLM-60-15-75 60.00V SLH-60-240-1800 60.00V

SLH-60-360-1800 60.00V

SLH-500-60-1800 500.00V

Table 2-8. VSense, Range II Gain Adjustment Settings (Item 8)

2-6

SLM/SLH Calibration Manual