Calibration

Elgar Electronics Corporation

2.6.2

CV M

ODE

, L

EVEL

H

IGH

, O

FFSET AND

G

AIN

, I

TEMS

31

AND

32

Set the DC power supply current limit to 0.1A.

1.

LOAD OFF.

2.

Press the

© key to go to Item 31.

3.

PRES ON.

4.

Set voltage to the V Input value shown in Table 2-31 for the model
being calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the DVM reading and the Voltage

Setting match.

V Input Voltage Setting

MODEL

MODEL

V Input Voltage Setting

SLM-60-30-150

61.5

0.800V SLH-60-120-600 61.5 0.800V

SLM-60-60-300

61.5

0.800V SLH-60-120-1200 61.5 0.800V

SLM-250-10-300

61.5

8.000V SLH-60-120-1800 61.5 0.800V

SLM-500-10-300

61.5

16.008V SLH-60-240-1200 61.5

0.800V

SLM-60-15-75

61.5

0.800V SLH-60-240-1800 61.5 0.800V

SLH-60-360-1800

61.5

0.800V

SLH-500-60-1800

61.5

16.008V

Table 2-31. CV Mode, Level High, Offset Adjustment Settings (Item 31)

8.

LOAD OFF.

9.

Press the

© key to go to Item 32.

2-28

SLM/SLH Calibration Manual