7 cp mode calibration, Cp m, Alibration – AMETEK SL Series Calibration Manual User Manual
Page 37

Elgar Electronics Corporation
Calibration
10.
PRES ON.
11.
Set voltage to the V Input value shown in Table 2-32 for the model
being calibrated.
12.
PRES OFF.
13.
LOAD ON.
14.
Adjust using the
keys until the DVM reading and the Voltage
Setting match.
15.
Repeat steps 1 through 14 (press
ª to return to Item 30) until no
further adjustments are necessary for the DVM reading and the
Voltage Setting to match.
MODEL
V Input Voltage Setting
MODEL
V Input Voltage Setting
SLM-60-30-150 61.5 60.000V SLH-60-120-600 61.5 60.000V
SLM-60-60-300 61.5 60.000V SLH-60-120-1200
61.5 60.000V
SLM-250-10-300 252
250.0V SLH-60-120-1800
61.5
60.000V
SLM-500-10-300 505
500.2V SLH-60-240-1200
61.5
60.000V
SLM-60-15-75 61.5 60.000V SLH-60-240-1800
61.5 60.000V
SLH-60-360-1800
61.5
60.000V
SLH-500-60-1800
505
500.2V
Table 2-32. CV Mode, Level High, Gain Adjustment Settings (Item 32)
2.7 CP
MODE
CALIBRATION
Make the connections shown in Figure 2-6.
Figure 2-6. Connections for CP Mode Calibration
SLM/SLH Calibration Manual
2-29