AMETEK SL Series Calibration Manual User Manual
Page 30
Calibration
Elgar Electronics Corporation
11.
Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-26 for the model being
calibrated.
12.
PRES ON.
13.
Set the current to the Current Setting value in Table 2-26 for the
model being calibrated.
14.
PRES OFF.
15.
LOAD ON.
16.
Adjust using the
keys until the DAM reading and the current
setting match.
17.
Repeat steps 1 through 16 (press
ª to return to Item 25) until no
further adjustments are necessary for the DAM reading and the
current setting to match.
MODEL Resistance
Setting
VSENSE Input Current Setting
SLM-60-30-150
2.003Ω 60.00V
30.00A
SLM-60-60-300
1.001Ω 60.00V
60.00A
SLM-250-10-300
25.00Ω 250.00V
10.000A
SLM-500-10-300
50.00Ω 500.00V
10.000A
SLM-60-15-75
4.000Ω 60.00V
15.000A
0.5000 Ω 60.00V
120.00A
SLH-60-120-600
SLH-60-120-1200
0.5000 Ω 60.00V
120.00A
SLH-60-120-1800
0.5000 Ω 60.00V
120.00A
SLH-60-240-1200
0.2500 Ω 60.00V
240.00A
SLH-60-240-1800
0.2500 Ω 60.00V
240.00A
SLH-60-360-1800
0.167 Ω 60.00V
359.281A
SLH-500-60-1800
8.3333Ω 500.00V 60.00A
Table 2-26. CR Mode, Level Low, Range II Gain Adjustment Settings (Item 26)
2-22
SLM/SLH Calibration Manual