Elgar Electronics Corporation

Calibration

10.

PRES ON.

11.

Set voltage to the V Input value shown in Table 2-30 for the model
being calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the DVM reading and the Voltage

Setting match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 29) until no

further adjustments are necessary for the DVM reading and the
Voltage Setting to match.

MODEL

V Input Voltage Setting

MODEL

V Input Voltage Setting

SLM-60-30-150 61.5V

60.000V

SLH-60-120-600 61.5V 60.000V

SLM-60-60-300 61.5V

60.000V

SLH-60-120-1200

61.5V 60.000V

SLM-250-10-300 252V

250.0V

SLH-60-120-1800

61.5V 60.000V

SLM-500-10-300 505V

500.2V

SLH-60-240-1200

61.5V 60.000V

SLM-60-15-75 61.5V 60.000V SLH-60-240-1800

61.5V 60.000V

SLH-60-360-1800

61.5V

60.000V

SLH-500-60-1800

505

500.2V

Table 2-30. CV Mode, Level Low, Gain Adjustment Settings (Item 30)

SLM/SLH Calibration Manual

2-27