beautypg.com

2 vsense, range ii offset and gain, items 7 and 8, Vsense, range ii offset and gain, items 7 and 8 -5 – AMETEK SL Series Calibration Manual User Manual

Page 13

background image

Elgar Electronics Corporation

Calibration

5.

Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-6 for the model being calibrated.

6.

Use the

…† keys to adjust the load DVM display until it matches the

DMM reading.

7.

Repeat steps 1 through 6 (press

ª to return to Item 5) until no further

adjustments are necessary for the load DVM to match the DMM
reading.

MODEL

MODEL

Input Voltage

Input Voltage

SLM-60-30-150 15.00V

SLH-60-120-600

15.00V

SLM-60-60-300 15.00V

SLH-60-120-1200

15.00V

SLM-250-10-300 30.00V SLH-60-120-1800

15.00V

SLM-500-10-300 60.00V SLH-60-240-1200

15.00V

SLM-60-15-75 15.00V

SLH-60-240-1800

15.00V

SLH-60-360-1800

15.00V

SLH-500-60-1800

60.00V

Table 2-6. VSENSE, Range I Gain Adjustment Settings (Item 6)

2.3.2

VS

ENSE

, R

ANGE

II O

FFSET AND

G

AIN

, I

TEMS

7

AND

8

1.

Press the

© key to go to Item 7.

2.

Set the DC Voltage Standard to the Input Voltage value shown in Table
2-7
for the model being calibrated.

3.

Use the

…† keys to adjust the load DVM display until it matches the

DMM reading.

MODEL

Input Voltage

MODEL

Input Voltage

SLM-60-30-150 0.200V

SLH-60-120-600 0.200V

SLM-60-60-300 0.200V

SLH-60-120-1200 0.200V

SLM-250-10-300 0.200V

SLH-60-120-1800 0.200V

SLM-500-10-300 0.400V

SLH-60-240-1200 0.200V

SLM-60-15-75 0.200V

SLH-60-240-1800

0.200V

SLH-60-360-1800

0.200V

SLH-500-60-1800

0.400V

Table 2-7. VSENSE, Range II Offset Adjustment Settings (Item 7)

4.

Press the

© key to go to Item 8.

SLM/SLH Calibration Manual

2-5