2 vsense, range ii offset and gain, items 7 and 8, Vsense, range ii offset and gain, items 7 and 8 -5 – AMETEK SL Series Calibration Manual User Manual
Page 13

Elgar Electronics Corporation
Calibration
5.
Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-6 for the model being calibrated.
6.
Use the
keys to adjust the load DVM display until it matches the
DMM reading.
7.
Repeat steps 1 through 6 (press
ª to return to Item 5) until no further
adjustments are necessary for the load DVM to match the DMM
reading.
MODEL
MODEL
Input Voltage
Input Voltage
SLM-60-30-150 15.00V
SLH-60-120-600
15.00V
SLM-60-60-300 15.00V
SLH-60-120-1200
15.00V
SLM-250-10-300 30.00V SLH-60-120-1800
15.00V
SLM-500-10-300 60.00V SLH-60-240-1200
15.00V
SLM-60-15-75 15.00V
SLH-60-240-1800
15.00V
SLH-60-360-1800
15.00V
SLH-500-60-1800
60.00V
Table 2-6. VSENSE, Range I Gain Adjustment Settings (Item 6)
2.3.2
VS
ENSE
, R
ANGE
II O
FFSET AND
G
AIN
, I
TEMS
7
AND
8
1.
Press the
© key to go to Item 7.
2.
Set the DC Voltage Standard to the Input Voltage value shown in Table
2-7 for the model being calibrated.
3.
Use the
keys to adjust the load DVM display until it matches the
DMM reading.
MODEL
Input Voltage
MODEL
Input Voltage
SLM-60-30-150 0.200V
SLH-60-120-600 0.200V
SLM-60-60-300 0.200V
SLH-60-120-1200 0.200V
SLM-250-10-300 0.200V
SLH-60-120-1800 0.200V
SLM-500-10-300 0.400V
SLH-60-240-1200 0.200V
SLM-60-15-75 0.200V
SLH-60-240-1800
0.200V
SLH-60-360-1800
0.200V
SLH-500-60-1800
0.400V
Table 2-7. VSENSE, Range II Offset Adjustment Settings (Item 7)
4.
Press the
© key to go to Item 8.
SLM/SLH Calibration Manual
2-5