Mode, Level low, offset and gain, items 29 and 30 -26 – AMETEK SL Series Calibration Manual User Manual
Page 34

Calibration
Elgar Electronics Corporation
2.6.1
CV
M
ODE
, L
EVEL
L
OW
, O
FFSET AND
G
AIN
, I
TEMS
29
AND
30
For CV Mode Calibration, set the DC power supply current limit to 0.1A.
1.
LOAD OFF.
2.
Press the
© key to go to Item 29.
3.
PRES ON.
4.
Set the power supply voltage to the V Input value shown in Table 2-29
for the model being calibrated.
5.
PRES OFF.
6.
LOAD ON.
7.
Adjust using the
keys until the DVM reading and the Voltage
Setting match.
MODEL
V Input Voltage Setting
MODEL
V Input Voltage Setting
SLM-60-30-150
61.5V 0.800V SLH-60-120-600 61.5V 0.800V
SLM-60-60-300
61.5V
0.800V SLH-60-120-1200
61.5V 0.800V
SLM-250-10-300
61.5V
8.000V SLH-60-120-1800
61.5V 0.800V
SLM-500-10-300
61.5V
16.008V SLH-60-240-1200
61.5V 0.800V
SLM-60-15-75
61.5V
0.800V SLH-60-240-1800
61.5V 0.800V
SLH-60-360-1800 61.5V 0.800V
SLH-500-60-1800
61.5V
16.008V
Table 2-29. CV Mode, Level Low, Offset Adjustment Settings (Item 29)
8.
LOAD OFF.
9.
Press the
© key to go to Item 30.
2-26
SLM/SLH Calibration Manual