Calibration

Elgar Electronics Corporation

2.6.1

CV

M

ODE

, L

EVEL

L

OW

, O

FFSET AND

G

AIN

, I

TEMS

29

AND

30

For CV Mode Calibration, set the DC power supply current limit to 0.1A.

1.

LOAD OFF.

2.

Press the

© key to go to Item 29.

3.

PRES ON.

4.

Set the power supply voltage to the V Input value shown in Table 2-29
for the model being calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the DVM reading and the Voltage

Setting match.

MODEL

V Input Voltage Setting

MODEL

V Input Voltage Setting

SLM-60-30-150

61.5V 0.800V SLH-60-120-600 61.5V 0.800V

SLM-60-60-300

61.5V

0.800V SLH-60-120-1200

61.5V 0.800V

SLM-250-10-300

61.5V

8.000V SLH-60-120-1800

61.5V 0.800V

SLM-500-10-300

61.5V

16.008V SLH-60-240-1200

61.5V 0.800V

SLM-60-15-75

61.5V

0.800V SLH-60-240-1800

61.5V 0.800V

SLH-60-360-1800 61.5V 0.800V

SLH-500-60-1800

61.5V

16.008V

Table 2-29. CV Mode, Level Low, Offset Adjustment Settings (Item 29)

8.

LOAD OFF.

9.

Press the

© key to go to Item 30.

2-26

SLM/SLH Calibration Manual