08/2013
10.7 Reset
You can reset (initialize) all sensor settings and sensor-independent settings 
separately from each other.
10.7.1 Resetting the measurement settings
pH
The following settings for pH measurements are reset to the default settings 
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required 
measured parameter in the measured value display and press the 
key.
ORP
The following settings for ORP measurements are reset to the default settings 
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required 
measured parameter in the measured value display and press the 
key.
The calibration data are reset to the default settings together with 
the measuring parameters. Recalibrate after performing a reset.
Setting
Default settings
Buffer YSI
Calibration interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
pH resolution
0.001
mV resolution
0.1
Asymmetry
0 mV
Slope
-59.2 mV
Man. temperature
25 °C
Single-point calibration
off
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C