beautypg.com

3 sample addition, Sample addi – YSI MultiLab 4010-3 User Manual

Page 62

background image

Ion concentration

MultiLab 4010-3

62

ba76143e03

08/2013

7.3.3 Sample addition

In the "Sample addition" procedure, a known amount of sample is added to the
standard solution.
The ion concentration in the sample is calculated from the change in potential.

Calibration

Measuring

1.

Select the measuring method (see section 7.3 S

ELECTING

THE

MEASU

-

RING

METHOD

, page 56).

2.

Prepare two standard calibration solutions.

3.

Perform a two-point calibration according to user guidance.

4.

The calibration record is displayed as soon as a stable value is
achieved for the second standard calibration solution.

01.08.2013 08:00

01.08.2013 08:00

#1 0.010 mg/l

20.2 mV 25.0 °C

#2 0.020 mg/l

79.2 mV 25.0 °C

Slp.: 54.1 mV
Sensor +++ (Ion type: Ag)

CALIBRATION
ADA 94/IDS BNC
B092500013

5.

Start the measurement with /[continue].
An entry window appears.

6.

Prepare the standard solution.

7.

If necessary, add the ISA/TISAB solution to the standard solution.

8.

Thoroughly rinse the electrode with deionized water.

9.

Immerse the electrode in the standard.

i

Immerse sensor in standard

Std. volume

100.0 ml

Std. conc.

1.0 mg/l

ISA/TISAB vol.

1.0 ml

continue

Sample addition

01.08.2013 08:00