YSI MultiLab 4010-3 User Manual
Page 76
Dissolved oxygen (D.O.)
MultiLab 4010-3
76
ba76143e03
08/2013
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
ProOBOD
calibration
evaluation
Menu item
Setting/func-
tion
Explanation
Calibration /
Calibration data
storage /Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using
played calibration record to
the USB-B (PC) interface or
the USB-A (USB printer)
interface .
Using
ibration records to the USB-B
(PC) interface or the USB-A
(USB printer) interface .
Quit the display with
Switch directly to the mea-
sured value display with
Calibration / Cali-
bration data stor-
age / Output to
USB flash drive or
printer
-
Outputs the stored calibration
data to the USB-A interface
(USB memory device/USB
printer)
Calibration /
Calibration data
storage / Output to
RS232/USB
-
Outputs the stored calibration
data to the USB-B interface (PC)
Display
Calibration record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94
or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92
or
S = 1.08 ... 1.10
Error
Error
S < 0.90
or
S > 1.10
Error elimination (see section 14 W