10 settings, 1 ph measurement settings, 1 settings for ph measurements – YSI MultiLab 4010-3 User Manual
Page 84: Section 10.1.1 s, Ettings, Measurements, Ee section 10.1.1 s

Settings
MultiLab 4010-3
84
ba76143e03
08/2013
10 Settings
10.1 pH measurement settings
10.1.1 Settings for pH measurements
The settings are made in the menu for calibration and measurement settings of 
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the 
After completing the settings, switch to the measured value display with 
Default settings are printed in bold.
Menu item
Possible 
setting
Explanation
Calibration /Calibration 
record
-
Displays the calibration record of the last calibration.
Calibration / Calibration 
data storage / Display
-
Displays the last calibration records (max. 10)
Calibration / 
Calibration data storage 
/ Output to RS232/USB
-
Outputs the stored calibration data to the USB-B interface 
(PC)
Calibration / Calibration 
data storage / Output to 
USB flash drive or 
printer
-
Outputs the stored calibration data to the USB-A interface 
(USB memory device/USB printer)
Calibration /Buffer
YSI
ConCal
NIST/DIN 
...
Buffer sets to be used for pH calibration.
More buffers and details: see section 10.1.2 B
Calibration /Single-point 
calibration
yes
no
Quick calibration with 1 buffer
Calibration /Calibration 
interval
1 ... 7 ... 999 
d
Calibration interval for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flash-
ing sensor symbol in the measuring screen.
Calibration /Unit for 
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of 
-59.2 mV/pH (100 x determined slope/Nernst slope).
QSC /First calibration
-
Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial cali-
bration was carried out with the connected IDS sensor.
QSC /Record of first cal-
ibration
-
Displays the calibration record of the QSC initial calibra-
tion.
QSC /Control calibra-
tion
-
Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was 
carried out with the connected IDS sensor.