Ion concentration

MultiLab 4010-3

60

ba76143e03

08/2013

9.

Immerse the electrode in the sample.

10.

Using <> <> and , select the values for the volume of
the sample (Sample volume) and the volume of the ISA/TISAB solution
(ISA/TISAB vol.).

11.

Select continue with <> <> and start the measurement with
.
When the measurement is finished an entry window appears.

12.

Add the standard (complexing agent or precipitating agent) to the sam-
ple.

13.

Using <> <> and , enter the values for the volume of the
standard solution (Std. volume) and concentration of the standard solu-
tion (Std. conc.) .

14.

If necessary, set the ion type for the ion in the standard (Ion type) with
<> <> and .
On selection of the ion type that can be defined (ION):
Using <> <> and , set the valence (Valency) and molar
weight (Molar mass) for the ion in the standard solution.

15.

Select continue with <> <> and start the measurement with
.
The measurement result is displayed when the measurement is com-
pleted.

16.

If necessary, start measuring further samples with .
Repeat steps 6 - 13 for all samples.

i

Add standard!

Ion type

S

Std. volume

1.0 ml

Std. conc.

1.0 mg/l

continue

Standard subtraction

01.08.2013 08:00

01.08.2013 08:00

Standard subtraction