3 dissolved oxygen, 4 conductivity, 3 dissolved oxygen 14.4 conductivity – YSI MultiLab 4010-3 User Manual
Page 118
What to do if...
MultiLab 4010-3
118
ba76143e03
08/2013
14.3 Dissolved oxygen
Error message,
OFL
Error message,
Error
14.4 Conductivity
Error message,
OFL
Error message,
Error
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the measur-
ing range
– Select a suitable measuring
medium
Cause
Remedy
– Sensor contaminated
– Clean the sensor
– Measured temperature value outside
the operating conditions
(display of OFL/UFL instead of a
temperature value)
– Keep to the temperature range
for the test sample
– Defective sensor
– Calibrate
– Change sensor cap
– Replace sensor
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the measur-
ing range
– Use suitable IDS conductivity
sensor
Cause
Remedy
– Sensor contaminated
– Clean the sensor and replace it
if necessary
– Calibration solution not suitable
– Check the calibration solutions