3 calibration interval – YSI MultiLab 4010-3 User Manual
Page 88
![background image](/manuals/716293/88/background.png)
Settings
MultiLab 4010-3
88
ba76143e03
08/2013
* Brand names or trade names are trademarks of their respective owners protected
by law.
10.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC
40).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
22
Reagecon 20 *
2.000
4.000
7.000
10.000
13.000
20 °C
23
Reagecon 25 *
2.000
4.000
7.000
10.000
13.000
25 °C
24
Chemsolute *
2.000
4.000
7.000
10.000
20 °C
25
USABlueBook
*
4.000
7.000
10.000
25 °C
The buffers are selected in the menu, pH /
/ Buffer (see section 10.1.1 S
84).
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
No.
Buffer set *
pH values
at
1.
Open the menu for measurement settings with
2.
In the Calibration / Calibration interval menu, set the calibration interval
with <><>.