Dissolved oxygen (D.O.)

ba76143e03

08/2013

75

8.2.6 Calibration data

Displaying the

calibration record

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the key in the measured value display.

8.

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

9.

Wait for the AutoRead measurement to be completed (audio signal).

When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.

When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
.
The calibration record is displayed and output to the interface.

10.

Switch to the measured value display with .

The calibration record is automatically transmitted to the interface
after calibrating.

01.08.2013 08:00

760 mmHg

25.1 °C