beautypg.com

2 measuring the temperature, 2 temperature compensation, Measuring the temperature – YSI MultiLab 4010-3 User Manual

Page 80: Temperature compensation

background image

Conductivity

MultiLab 4010-3

80

ba76143e03

08/2013

9.1.2 Measuring the temperature

For reproducible conductivity measurements, it is essential to measure the
temperature of the test sample.
IDS sensors measure the temperature with a temperature sensor integrated in
the IDS sensor.

9.2

Temperature compensation

The calculation of the temperature compensation is based on the preset refer-
ence temperature, 20 °C or 25 °C. It appears on the display as Tr20 or Tr25.

You can select one of the following temperature compensation methods:

Nonlinear temperature compensation (nLF) according to ISO 7888

Linear temperature compensation (lin) with adjustable coefficients of
0.000 ... 3.000 %

/

K

No temperature compensation (off)

Application tips

Select the following temperature compensations given in the table according to
the respective test sample:

The reference temperature and temperature compensation are set
in the menu for the parameter, conductivity (see section 10.5.1 S

ET

-

TINGS

FOR

IDS

CONDUCTIVITY

SENSORS

, page 92).

Test sample

Temperature compensation

Display

Natural water
(ground water, surface
water, drinking water)

nLF
according to ISO 7888

nLF

Ultrapure water

nLF
according to ISO 7888

nLF

Other aqueous solu-
tions

lin Set linear
temperature coefficient
0.000 ... 10.000 %/K

lin

Salinity (seawater)

Automatic nLF according to
IOT (International Oceano-
graphic Tables)

Sal, nLF