YSI MultiLab 4010-3 User Manual
Page 52
Ion concentration
MultiLab 4010-3
52
ba76143e03
08/2013
Continuing with
three- to seven-point
calibration
Repeat the steps 8 to 13 in the same way with the third and further standard
solutions as necessary. The new calibration values are displayed after the last
calibration step was completed.
15.
When calibrating without temperature sensor:
Measure the temperature of the standard solution using a thermom-
eter.
Use
⇅ ] to select the setting of the temperature.
Use <> <> to set the temperature.
Use
⇅ ] to select the setting of the concentration.
16.
Set the concentration of the standard solution with <> <> and
press
The standard solution is measured.
The measured value is checked for stability (AutoRead).
17.
Wait for the end of the AutoRead measurement or accept the calibra-
tion value with
The calibration display for the next standard solution appears.
18.
Press
or
Finish the calibration procedure as a two-point calibration with
The new calibration values are displayed.
01.08.2013 08:00
01.08.2013 08:00
Standard
Based on the calibration data, the calibration curve is determined
in sections, according to the Nernst equation modified by Nikolski.